Show simple item record

dc.contributor.authorIping Petterson, IE
dc.contributor.authorDay, JC
dc.contributor.authorFullwood, LM
dc.contributor.authorGardner, B
dc.contributor.authorStone, N
dc.date.accessioned2016-05-05T09:01:24Z
dc.date.issued2015-09-28
dc.description.abstractWe demonstrate the first use of a multifibre Raman probe that fits inside the bore of a hypodermic needle. A Raman probe containing multiple collection fibres provides improved signal collection efficiency in biological samples compared with a previous two-fibre design. Furthermore, probe performance (signal-to-noise ratios) compared favourably with the performance achieved in previous Raman microscope experiments able to distinguish between benign lymph nodes, primary malignancies in lymph nodes and secondary malignancies in lymph nodes. The experimental measurements presented here give an indication of the sampling volume of the Raman needle probe in lymphoid tissues. Liquid tissue phantoms were used that contained scattering medium encompassing a range of scattering properties similar to those of a variety of tissue types, including lymph node tissues. To validate the appropriateness of the phantoms, the sampling depth of the probe was also measured in excised lymph node tissue. More than 50 % of Raman photons collected were found to originate from between the tip of the needle and a depth of 500 μm into the tissue. The needle probe presented here achieves spectral quality comparable to that in numerous studies previously demonstrating Raman disease discrimination. It is expected that this approach could achieve targeted subcutaneous tissue measurements and be viable for use for the in vivo Raman diagnostics of solid organs located within a few centimetres below the skin's surface. Graphical Abstract Schematic of multi-fibre Raman needle probe with disposible tips and proximal optical filtration.en_GB
dc.description.sponsorshipThis work is funded by a UK National Institute for Health Research (NIHR) Invention for Innovation (i4i) grant number II_LA_1111_20007.en_GB
dc.identifier.citationAnalytical and Bioanalytical Chemistry, 2015, Vol. 407, Issue 27, pp. 8311 - 8320en_GB
dc.identifier.doi10.1007/s00216-015-9021-7
dc.identifier.urihttp://hdl.handle.net/10871/21388
dc.language.isoenen_GB
dc.publisherSpringer Verlagen_GB
dc.relation.urlhttp://www.ncbi.nlm.nih.gov/pubmed/26416020en_GB
dc.rights.embargoreasonPublisher policyen_GB
dc.rightsThis is the author accepted manuscript. The final version is available from Springer Verlag via the DOI in this record.en_GB
dc.subjectDiagnosticsen_GB
dc.subjectLymph nodesen_GB
dc.subjectNeedleen_GB
dc.subjectNon-invasiveen_GB
dc.subjectRaman spectroscopyen_GB
dc.titleCharacterisation of a fibre optic Raman probe within a hypodermic needle.en_GB
dc.typeArticleen_GB
dc.identifier.issn1618-2642
exeter.place-of-publicationGermany
dc.identifier.journalAnalytical and Bioanalytical Chemistryen_GB


Files in this item

This item appears in the following Collection(s)

Show simple item record